Fail-Safe Architecture for Closed-Loop Cycle-Ergometer Cardiac Rehabilitation Based on BLE Sensors and Smart Trainers

Abstract

A fail-safe architecture for closed-loop cardiac rehabilitation on a cycle ergometer with BLE devices is proposed, where only a safety kernel controls the trainer. It is verified using emulators, a physiological patient model and fault injection. Experiments confirmed time-bounded transition to a safe state and revealed limits of reactive protection

Author (co-authors)
First name Last name Institutional affiliation E-mail Phone number ORCID ID Academic status, position Institution address Author contribution(s) Institutional affiliation
Ivan
Boichuk
ivan.v.boichuk@lpnu.ua
0009-0005-6141-3161
PhD student and Assistant at the Software Department
S. Bandery street 12, Lviv, 79013, Ukraine
Investigation
Methodology
Software
Validation
Visualization
Writing – Original Draft Preparation
Lviv Politechnic National University
Dmytro
Fedasyuk
dmytro.v.fedasyuk@lpnu.ua
0000-0003-3552-7454
Doctor of Technical Sciences (Sc.D.), Professor
S. Bandery street 12, Lviv, 79013, Ukraine
Project Administration
Supervision
Writing – Review & Editing
Lviv Politechnic National University